Replacement of the UV Lamp on Millipore Systems MILLI-Q  Century (Mfg after mid 2002)

Complete list of replacement
Lista completa di ricambi per Millipore
Komplette Millipore ersatzteil liste

QUANTUM® comparable replacement
QGARD1® comparable Replacement
QGARD2® comparable Replacement
QPAK1® comparable Replacement
QPAK2® comparable Replacement
PROGARD1® camparable replacement
PROGARD2® comparable replacement
PROGTL1CL1 comparable replacement
SYMPAK2®  comparable replacement
MILLIPAK® comparable replacement
UV Lamps for MilliQ® 1996>07/2002
UV Lamps for MilliQ® Century® 2002>



UV LAMP No. UV0002 (for MilliQ mfg after mid 2002)

item N. UV0002
Replacement Instructions for UV Lamp ("Century" Range MiIIi-Q)
The purpose of this document is to provide instructions on how to replace a UV Lamp installed on
MiIli-Q Water Purification Systems. These instructions apply to a UV Lamp with NOION AQUA CatalogueNumber UV 0002
(Millipore No.ZMQUVLPO1).

Replacement of the UV Lamp
1. Place the System into STANDBY Mode. Turn off the electrical power to the system.
2. Remove the System cover.
3. Locate the UV Lamp and its Stainless Steel Housing. See the picture for its replacement.
4. Remove the UV Ballast Wires that plug into the top of the UV Lamp.
5. Remove the UV lamp from the system. DO NOT remove the Quartz Sleeve (the glass piece) located
    between the actual lamp and the inside of the Stainless Steel Housing.
6. lnstall the replacement (new) UV Lamp by reversing Steps 5 and 4 above.
Put the system cover back on.
8. Power on the System.


UV lamp lifetime reset:
In STANDBY or PRE-OPERATE mode Press:
"
Measure + Cleaning + Menu" buttons together to enter the Service Mode
Press 4 times on MENU button (SERVICE: RESET UV LAMP)
Press
MEASURE + CLEANING buttons together to confirm reset
The UV lamp timer is reset for 365 days
Press on
MENU button for 2 seconds to return to start position
The replacement procedure is now completed. The Milli-Q is now ready for use.

Order
remove lamp

Istruzioni / How to / Betriebsanleitungen

Millipore Water systems:

Modo d'uso del MilliQ
Produrre una quantità precisa di acqua pura
Come sostituire le cartucce del MilliQ

How to change the cartridges
How to change the UV lamp UV0001